A strain gauge (also spelled strain gage) is a bonded resistance sensor that converts mechanical surface elongation into a proportional change in electrical resistance. Because a standard 120-ohm metallic foil gauge has a gauge factor slightly above 2, its tiny resistance shift requires a 5 V or 12 V Wheatstone bridge to produce readable millivolt outputs.
This site (Maximizing Value at NI) is independent of National Instruments and Emerson. Related guides in Sensors and DAQ cover strain gauge vs. load cell selection, quarter, half, and full Wheatstone bridge wiring, and three-wire quarter-bridge checks.
| Parameter | Verified baseline | Engineering significance |
|---|---|---|
| Measurand | Average unit elongation (ÎL/L), dimensionless microstrain | Measures local surface deformation over a nominal gauge length |
| Nominal foil resistance | 120 Ω (standard metal foil) | Low-resistance 120 Ω gauges are less prone to humidity-induced error than higher-resistance gauges |
| Fractional resistance shift | 2 Ă 10â»â¶ to 2 Ă 10â»Âł on a 120 Ω gauge | Too small by 3 to 6 orders of magnitude for a standard ohmmeter |
| Metallic foil gauge factor | Slightly above 2 | Transfer function linking resistance change to applied strain |
| Foil active area and stretch | 2â10 mmÂČ; up to at least 10% with proper bonding | General-purpose sensing across metallic test structures |
| Bridge excitation and output | 5 V or 12 V input; millivolt output | Balanced bridge turns small resistance shifts into measurable voltage |
| Recommended operating limit | ±3000 microstrain | Exceeding the design limit causes unrecoverable degradation |
What is a strain gauge and how does it differ from an extensometer or load cell?
Invented independently by Edward E. Simmons and Arthur C. Ruge in 1938, a strain gauge (also spelled strain gage) most commonly uses an insulating flexible backing supporting a metallic foil pattern bonded directly to a test surface, as documented in Wikipediaâs strain gauge reference. Unlike an extensometer or calibrated load cell, a bonded gauge reports local surface elongation rather than total displacement or calibrated force.
Under ASTM E251, metallic bonded resistance strain gages differ from extensometers because they measure average unit elongation (ÎL/L) over a nominal gauge length rather than total elongation between definite gauge points. Strain gages offer essentially infinite resolution, though their electrical response to strain is low and requires well-characterized performance parameters to yield meaningful field data.
ASTM E251 defines uniform test methods for five strain gage performance characteristics, and highlights a critical metrology distinction. Only resistance measurement is non-destructive and can be performed on each individual gauge; mechanical performance tests are destructive, so manufacturers treat those results statistically and apply them to the remaining lot or batch. Consequently, individual bonded strain gages cannot be directly calibrated, and where calibration and traceability to a standard are required, individual strain gages should not be employed.
ASTM E251 explicitly excludes complete transducers such as load cells and extensometers that incorporate bonded resistance strain gages as sensing elements. Legal metrology load cells instead fall under frameworks such as OIML R 60, which ranks load cells by overall performance into four accuracy classes: Class A, Class B, Class C, and Class D. While strain gauge technology was a primary focus in developing OIML R 60, the standard also accommodates hydraulic, pneumatic, vibratory frequency, and magnetic force principles. Beyond weighing and force transducers, strain gauge technology is widely used in pressure sensors built with silicon, polysilicon, metal film, thick film, or bonded foil elements. For deciding between bare gauges and packaged force transducers, see our strain gauge vs. load cell comparison.
How does a strain gauge convert mechanical deformation into an electrical signal?
When a bonded foil grid stretches or compresses with the test surface, its electrical resistance shifts by a tiny fractionâapproximately 2 Ă 10â»â¶ to 2 Ă 10â»Âł on a standard 120-ohm gaugeâwhich is 3 to 6 orders of magnitude too small for a simple ohmmeter. Wiring the gauge into a balanced Wheatstone bridge driven at 5 V or 12 V converts that shift into a millivolt signal.
As explained in Penn State University strain gauge lecture notes, strain is dimensionless, and microstrain is a dimensionless unit used to express small surface deformations cleanly. In metallic test parts, axial loading also produces lateral contraction: UTRGV strain measurement lecture notes note that Poissonâs ratio for most metals is approximately 0.3, meaning transverse strain is about -30% of the axial strain.
Because standard metal foil strain gages commonly have a nominal unstrained resistance of 120 ohms, direct resistance measurement with an ordinary ohmmeter lacks the resolution needed to detect fractional changes spanning 2 Ă 10â»â¶ to 2 Ă 10â»Âł. Engineers therefore rely on the Wheatstone bridgeâinvented by Samuel Hunter Christie in 1833 and improved and popularized by Sir Charles Wheatstone in 1843, according to Wikipediaâs Wheatstone bridge article. When all four bridge resistors are identical (R1 = R2 = R3 = R4), the bridge is balanced and the output voltage Vo equals 0. Applying a typical excitation voltage of 5 V or 12 V to the input leads produces differential output readings in millivolts whenever mechanical strain unbalances the active gauge arm.
What is gauge factor and how do foil and semiconductor gauges compare?
Under ASTM E251, gage factor is the transfer function relating fractional resistance change to applied mechanical strain, and strain measurement accuracy can be no better than the accuracy of the gage factor. Common metallic foil gauges have a gauge factor slightly above 2, whereas semiconductor piezoresistors offer a larger gauge factor for small strains at the cost of higher fragility and temperature sensitivity.
Foil gauges are the general-purpose choice for structural testing. They typically have active areas of about 2â10 mmÂČ and, with careful installation, the right gauge, and the right adhesive, can measure strains up to at least 10%. Their modest gauge factor requires sensitive bridge conditioning, but their mechanical robustness and predictable alloy behavior make them standard across experimental stress analysis.
Semiconductor strain gaugesâoften called piezoresistorsâare frequently chosen when measuring very small strains because their gauge factor is larger than that of metallic foil gauges. However, semiconductor gauges are more expensive, more sensitive to temperature changes, and more fragile than foil gauges. For deeper mathematical treatment of lot calibration and sensitivity corrections, review our guide to strain gauge factor and transverse sensitivity.
| Attribute | Metallic foil strain gauge | Semiconductor (piezoresistive) strain gauge |
|---|---|---|
| Sensing element | Metallic foil grid on an insulating flexible backing | Semiconductor piezoresistive element |
| Typical gauge factor | Slightly above 2 | Larger than metallic foil gauges |
| Active area and range | About 2â10 mmÂČ; up to at least 10% strain with proper bonding | Preferred for small-strain measurements |
| Temperature sensitivity | Mitigated with self-temperature-compensated constantan or Karma alloys | More sensitive to temperature changes |
| Mechanical durability and cost | Rugged general-purpose construction | More fragile and more expensive than foil gauges |
When should you use a quarter, half, or full Wheatstone bridge?
Bridge selection depends on how many active strain gauges you can bond to the structure: a quarter bridge uses one active gauge and three fixed resistors, a half bridge uses two active gauges and two fixed resistors, and a full bridge uses four active gauges. Adding active gauges raises bridge output and aids thermal rejection, while 3-wire or 4-wire Kelvin wiring cancels lead-wire temperature drift.
In UTRGVâs strain measurement classification, the three standard bridge arrangements trade wiring complexity against sensitivity and compensation. All three networks are commonly driven with 5 V or 12 V excitation and read in millivolts across the output terminals. Whenever long cable runs experience ambient temperature swings, lead-wire resistance changes can masquerade as strain; cancelling those lead-wire temperature effects requires either a 3-wire bridge or a 4-wire ohm circuit (also called a 4-wire Kelvin connection).
| Bridge configuration | Active strain gauges | Fixed resistors | Lead-wire compensation option |
|---|---|---|---|
| Quarter bridge | One active strain gauge | Three fixed resistors | 3-wire bridge or 4-wire Kelvin connection |
| Half bridge | Two active strain gauges | Two fixed resistors | 3-wire bridge or 4-wire Kelvin connection |
| Full bridge | Four active strain gauges | None | Balanced four-arm bridge wiring |
For detailed arm placement and verification steps, see our quarter, half, and full Wheatstone bridge comparison and three-wire quarter-bridge checks.
What strain gauge types, rosettes, and non-contact alternatives exist?
Beyond single-axis linear foil gauges, multi-axis rosettesâincluding 90° rosettes with 2 measuring directions and 45° rosettes with 3 measuring directionsâcapture biaxial or unknown principal strain fields. When bonded electrical gauges are impractical, fiber optic sensing measures point or distributed strain along an optical fiber, and digital image correlation (DIC) maps full-field surface strain without physical contact.
Choosing the right grid pattern starts with whether you know the principal strain directions on the test part. A single-axis linear gauge aligns with one known loading direction. A 90° rosette combines two perpendicular grids for biaxial strain states, while a 45° rosette combines three measuring directions so engineers can resolve principal strains and angles when directions are unknown in advance. To work through principal strain reduction, see our strain gauge rosette principal strain and orientation guide.
Optical methods extend strain measurement beyond bonded foil limits. Fiber optic sensing measures strain either at predetermined points or distributed continuously along an optical fiber. Digital image correlation (DIC) uses one or two cameras paired with tracking software to monitor surface features without contacting the specimen.
| Measurement approach | Sensing directions or coverage | Contact method | Best-fit engineering scenario |
|---|---|---|---|
| Linear foil gauge | Single axis | Bonded resistance grid | Known uniaxial strain direction on structural parts |
| 90° rosette | 2 measuring directions | Bonded multi-grid foil | Biaxial strain states along orthogonal axes |
| 45° rosette | 3 measuring directions | Bonded multi-grid foil | Multi-axis stress analysis when principal directions are unknown |
| Fiber optic sensing | Point or distributed along fiber | Optical fiber sensor | Long structures requiring distributed or point optical sensing |
| Digital image correlation (DIC) | Full-surface camera field | Non-contact (one or two cameras) | Non-contact surface deformation mapping with tracking software |
What causes strain gauge measurement errors and how do you avoid them?
Field accuracy depends heavily on controlling additive thermal output, transverse sensitivity, overloading, humidity-induced leakage, and bonding quality. Under ASTM E251, thermal output can exceed the structural load signal itself, while poor surface preparation, adhesive choice, protective coating, installer skill, or instrumentation design can allow error signals to swamp true structural deformations.
Thermal output and alloy or dummy-gauge compensation
ASTM E251 identifies thermal output as an additive error driven by temperature changes rather than mechanical load. Because it can exceed the load-induced signal, thermal output must be characterized on specimens of the same test material. Most metallic foil gauges use constantan or Karma alloys engineered for self-temperature compensation so that thermal resistance changes largely cancel the thermal expansion mismatch of the bonded substrate. When using non-self-temperature-compensated gauges such as isoelastic alloy, engineers wire an identical dummy gauge bonded to an unstrained sample of the same material into an adjacent bridge arm so thermal shifts cancel.
Transverse sensitivity, overloading, humidity, and bonding
Transverse sensitivity describes a gaugeâs response to strain perpendicular to its primary axis. ASTM E251 notes that although transverse sensitivity is usually much less than 10% of the gage factor, large errors can still occur if the value is not known with reasonable precision.
Mechanical and environmental limits are equally strict. Overloading a strain gauge beyond its design microstrain limit causes unrecoverable performance degradation, so good engineering practice avoids stressing standard gauges beyond ±3000 microstrain. In humid environments, unprotected lead wires can corrode and form parasitic resistance paths; low-resistance 120-ohm strain gauges are less prone to this humidity-induced error than higher-resistance gauges, and protective coatings are essential outside dry laboratory settings. Finally, adhesive selection must match test duration: cyanoacrylate glue is appropriate for short-term installations lasting up to some weeks, whereas epoxy glue is required for long-lasting installations, paired with rigorous surface cleaning and skilled bonding.
Strain gauge selection and installation checklist
| Control item | Verified engineering requirement | Primary risk prevented |
|---|---|---|
| Calibration and traceability | Use lot-based ASTM E251 parameters; switch to an OIML R 60 load cell if direct calibration is required | Relying on an uncalibrated individual gauge where traceability is required |
| Resistance and humidity protection | Use standard 120 Ω foil gauges and protective coatings on leads in humid environments | Parasitic leakage resistance and lead-wire corrosion |
| Alloy and thermal compensation | Select self-temperature-compensated constantan or Karma alloy, or add a matched dummy gauge | Additive thermal output swamping the structural load signal |
| Bridge and lead wiring | Match quarter, half, or full bridge to active gauge count; add 3-wire or 4-wire Kelvin wiring | Lead-wire resistance drift during ambient temperature swings |
| Adhesive selection | Use cyanoacrylate for short-term tests (up to some weeks) or epoxy for long-lasting installations | Bond creep, hysteresis, and premature delamination |
| Strain operating envelope | Keep standard bonded gauges within ±3000 microstrain | Unrecoverable overload degradation |
Sources
- ASTM E251 Standard Test Methods for Performance Characteristics of Metallic Bonded Resistance Strain Gages â ASTM International
- OIML R 60 Metrological regulation for load cells â International Organization of Legal Metrology (OIML)
- Stress, Strain, and Strain Gages (John M. Cimbala) â Penn State University
- Strain Gage Lab: Measurement of Strain â Penn State University
- Strain Measurements (Isaac Choutapalli) â University of Texas Rio Grande Valley
- Strain gauge â Wikipedia
- Wheatstone bridge â Wikipedia



